Armstrong Steam QM-3 automatic steam quality monitor. Simultaneously measures steam dryness, superheat, and non-condensable gases (NCGs). EN 285 compliant with RS485/Modbus data logging.

The Armstrong Steam QM-3 is an automatic steam quality monitor intended to replace manual testing of pure steam and provide real-time data proving that steam quality meets applicable requirements. It performs three critical tests: calculating dryness, calculating superheat, and quantifying non-condensable gases (NCG). The NCG measurement is performed first; if it is within range, dryness and superheat measurements will be performed.

This European standard specifies requirements and the relevant tests for large steam sterilizers, primarily used for the sterilization of medical devices and their accessories but has been adopted by most pharmaceutical industries.
The patented Steam Quality Monitor QM-3 is an alternative procedure to the one described in EN285 that has been shown to give equivalent results to the method specified in this European Standard.
Until now steam quality measurement has been a time-intensive, unreliable and potentially unsafe process. Steam QM-3 is not only more reliable and safer than manual testing, the unit is also portable, so it can be easily transported to multiple points on your steam line.
| Feature | Manual Method | Automatic Method (Steam QM-3) |
|---|---|---|
| Description | A sample of clean steam is condensed, and the enthalpy allows measurement of steam dryness, and NCG content. Temperature measurement before condensation identifies an excess of superheat temperature. | Reducing steam pressure to atmosphere allows measurement of steam dryness. Steam temperature and pressure measurements detect superheat. Volume of non-condensable gases are compared to condensate. |
| Disadvantages / Advantages | Time Consuming: Typically requires two people, and can take up to three hours per measurement point. Trending: Impossible to monitor a trend over time. Unsafe: Inherent safety risks involved in sampling live steam. Unreliable: Results depend on the skill of the technician. | Quick and Easy: Simple to install. Trending: Continuous measurements provide trending data over time. Safe: Installed while the steam valve is closed, making it much safer. Reliable: Reliable and accurate within +/- 1% of steam dryness. |
| Parameter | Specification |
|---|---|
| Calorimeter operating temperature range (T1) | 0 – 150 °C (32 – 302 °F) |
| Maximum allowable calorimeter temperature (T1) | 180 °C (356 °F) |
| Calorimeter operating saturated steam pressure range (P) | 0.5 – 4 barg (7 – 60 psig) |
| Superheated steam operating temperature range (T2) | 0 – 180 °C (32 – 356 °F) |
| Dryness fraction | 0.85 – 1 |
| Dryness fraction accuracy | ± 0.01 of display |
| Non-condensable gases content | ≤ 15% |
| Condensate temperature range (T3) | 0 – 80 °C (32 – 176 °F) |
| Operating condensate temperature (T3) | 65 °C (149 °F) |
| Estimated steam consumption | 1.5 kg/h (3.3 lbs/h) @ 3 barg (45 psig) |
| Estimated water consumption | 15 L/h (4 gal/h) @ 10 °C (50 °F) |
| Electrical | 115/230 VAC (±10%), 50/60 Hz, 100 W |
| Dimension | in | mm |
|---|---|---|
| A – Cabinet Width | 22 | 550 |
| B – Width | 24 | 600 |
| C – Cabinet Height | 40 | 1000 |
| D – Depth | 7.5 | 190 |
| E – Width | 17 | 430 |
| F – Height | 37 | 945 |
| G - Height from inlet tee to bottom | 26 | 660 |
| H - Height from cabinet steam inlet to bottom | 7.5 | 190 |
| Cabinet Weight | 42 lb | 19 kg |
| Total Weight | 55 lb | 25 kg |

Calorimeter Assembly Components:

Cabinet Interior Components:
Burn Hazard
Uninsulated components upstream of cabinet may be hot. Do not touch when unit is working. Allow to cool before moving or servicing unit. Live steam will cause burns; condensate water may cause them. Skin exposure to 140 °F (60 °C) water for only five seconds may cause a second degree burn.
Shock Hazard
High voltages present inside equipment. Electrical installation must be performed by qualified personnel. Disconnect power before performing any electrical service.


| Description | Part Nr. |
|---|---|
| Steam Quality Monitor QM-3 Package | D44212 |
| Insulation jackets (2pcs) | D43805 |
| Stainless Steel Wall mount | D44160 |
| Calorimeter assembly | D49150 |
| Condenser with fittings | D79386 |
| Main board with display | D44124 |
| Relay board | D44125 |
| Vessel Non Condensable Gases assembly with expansion coil | D44126 |
| Vessel Condensate flow meter assembly with fittings | D46738 |
| JUMO PT100 (T1 and T2) | D44110 |
| JUMO Pressure transmitter | D44117 |
| Heating element | D44118 |
| BURKERT 3/2 solenoid valve (EV0) | D44119 |
| BURKERT 2/2 solenoid valve (EV1 and EV3) | D44120 |
| BURKERT 3/2 solenoid valve (EV2) | D44121 |
| JUMO PT100 (T3) | D44122 |
| JUMO Differential pressure sensor | D44123 |
| Orifice Plate with Gasket | D40020 |
| Gasket PTFE for 1/2'' flexible hose (bag with 3 pieces) | D53335 |
A 9 ft (3 m) cord is provided with an M12 connector for Steam QM-3 cabinet connection. The logger end of the cord must be wired by the customer. Program the logger with the following information:
Data are sent in different registers using Modbus function 16 (0x10). All values are SI units. Decimal values are not used (e.g., 19.2 will show as 192).
| Register | Name | Factor | Unit | Data Type |
|---|---|---|---|---|
| 40 001 | T1 | x 10 | °C | (Integer value) |
| 40 002 | T2 | x 10 | °C | (Integer value) |
| 40 004 | P1 | x 10 | bar[a] | (Integer value) |
| 40 007 | X | x 1000 | N/A | (Integer value) |
| 40 008 | Q | x 10 | kg/h | (Integer value) |
| 40 011 | % NCG | x 10 | N/A | (Integer value) |
| 40 012 | Alarms | N/A | N/A | Bit 1: Alarm 1, Bit 2: Alarm 2, Bit 3: Alarm 3, Bit 4: Default 4, Bit 5: Default 5, Bit 6: Default 6 |
In slave mode, you can send queries with function 0x02 and 0x04. The length of the query and response must remain less than 23 bytes (max 3 registers per query).
Function 0x02 Read Discrete Inputs:
| Register | Name | Factor | Unit | Description |
|---|---|---|---|---|
| 10001 | Alarm1 | N/A | 0: disable - 1: enable | Toggle to 1 when the alarm 1 is on |
| 10002 | Alarm2 | N/A | 0: disable - 1: enable | Toggle to 1 when the alarm 2 is on |
| 10003 | Alarm3 | N/A | 0: disable - 1: enable | Toggle to 1 when the alarm 3 is on |
| 10004 | Default4 | N/A | 0: disable - 1: enable | Toggle to 1 when the default 4 is on |
| 10005 | Default5 | N/A | 0: disable - 1: enable | Toggle to 1 when the default 5 is on |
| 10006 | Default6 | N/A | 0: disable - 1: enable | Toggle to 1 when the default 6 is on |
Function 0x04 Read Inputs Registers:
| Register | Name | Factor | Unit | Description |
|---|---|---|---|---|
| 30001 | T1 | 10x | °C | Integer (16 bits) |
| 30002 | T2 | 10x | °C | Integer (16 bits) |
| 30004 | Alarm3 | 10x | Bar [a] | Integer (16 bits) |
| 30007 | Default4 | 1000x | - | Integer (16 bits) |
| 30008 | Default5 | 10x | Kg/hr | Integer (16 bits) |
| 30011 | Default6 | 10x | - | Integer (16 bits) |